Gilles Zambardi
AST Expert
BIOMERIEUX
LA BALME LES GROTTES, Rhone-Alpes, France
Disclosure: biomerieux (Employee)
Marion Pompilio
Technician in Microbiology
BIOMERIEUX
La Balme Les Grottes, Rhone-Alpes, France
Disclosure: BioMérieux (Employee)
EA | CA | Very Major Error Rate | Major Error Rate | Minor Error Rate |
99.0% (592/598) | 92.8% (555/598) | 1.9% (1/53) | 0.0% (0/478) | 7.0% (42/598) |
MIC/Interp µg/ml | ≤0.016 S | 0.032 S | 0.064 S | 0.125 S | 0.25 S | 0.5 S | 1 S | 2 S | 4 I | 8 R | 16 R | 32 R | 64 R | 128 R | ≥256R |
BMD | 0 | 0 | 0 | 5 | 122 | 172 | 98 | 81 | 67 | 15 | 7 | 1 | 0 | 1 | 29 |
ETEST® PLZ | 0 | 0 | 0 | 5 | 59 | 232 | 93 | 94 | 65 | 11 | 7 | 3 | 0 | 0 | 30 |